MIMIZUKU Spectifications

Overview

Wavelength Coverage2 to 38 micronwith 3 channels
Size2m x 2m x 2m
Weigth~ 2.3 ton

NIR Channel

Wavelength Coverage2 to 5.6 micron
DetectorInSb Alladin II 1024x1024
Pixel Scale0.11"/pix
Field of View2'x2'w/o Field Stacker
1'x2' x 2fieldw/ Field Stacker
Imaging FiltersK, Ldash, M+ J,H (optional)
Spectroscopy2.8-5.5um R~180Si-grism
Slit0.3" x 60"

MIR-S Channel

Wavelength Coverage6 to 26 micron
DetectorSi:As AQUARIUS 1024x1024
Pixel Scale0.11"/pix
Field of View2'x2'w/o Field Stacker
1'x2' x 2fieldw/ Field Stacker
Imaging Filtersmedium-band filters (Δλ~1um)
narrow-band filters (Δλ~0.2um)
Spectroscopy7.5-13.5um R~230Ge-grism
16.5-25.5um R~160Si-grism
Slit0.3"or0.6" x 60"
Sensitivity 1sig1sec50mJy10um med-band imaging
130mJy20um med-band imaging
250mJy10um spectroscopy
650mJy20um spectroscopy

MIR-L Channel

Wavelength Coverage24 to 36 micron
DetectorSi:Sb 128x128or Si:Sb 1024x1024 (optional
Pixel Scale0.44"/pix
Field of View56"x56"w/o Field Stacker
28"x56" x 2fieldw/ Field Stacker
Imaging Filtersmedium-band filters (Δλ~3um)
Spectroscopy26-38um R~60)Optional
Slit1.2" x 50"
Sensitivity 1sig1sec200mJy30um med-band imaging
400mJy38um med-band imaging

Related Documents